36 Specifications
408
Measurement Specifications
See the Appendix on Default Settings for a list of the settings the monitor is initially shipped with.
ECG/Arrhythmia/ST/QT
Complies with IEC 60601-2-25:1993 + A1:1999 /EN60601-2-25:1995 + A1:1999, IEC 60601-2-
27:2005/EN60601-2-27:2006, IEC 60601-2-51:2003 /EN 60601-2-51:2003 and AAMI EC11/
EC13:1991/2002.
Display Specifications
19" WSXGA Type 482 mm active matrix color LCD
(TFT)
Resolution 1680 x 1050 (WSXGA+)
Refresh frequency 58 Hz
Useful screen 409,5 mm x 255.9 mm
Pixel size 0.244 mm x 0.244 mm
ECG/Arrhythmia/ST Performance Specifications
Cardiotach Range Adult/pedi: 15 to 300 bpm
Neo range: 15 to 350 bpm
Accuracy ±1% of range
Resolution 1 bpm
Sensitivity 200 µV
peak
PVC Rate Range 0 to 300 bpm
Resolution 1 bpm
ST Numeric Range -20 to +20 mm
Accuracy ±0.5 mm or 15%, whichever is greater
Resolution 0.1 mm
QT Numeric Range 200 to 800 ms
Accuracy ±30 ms
Resolution 8 ms
QTc Numeric Range 200 to 800 ms
Resolution 1 ms
ΔQTc Numeric Range -600 to +600 ms
Resolution 1 ms
QT-HR Numeric Range - adult 15 to 300 bpm
Range - pediatric and
neonatal
15 to 350 bpm