36 Specifications
407
1
The short range radio interface is compatible with the following telemetry devices: TRx4841A/TRx4851A Intellivue
Telemetry System Transceiver
ECG Sync Pulse
Output (rectangular
pulse)
Cable detection Yes
Marker In No
Wave Output No
Connector Binder Series 709/719
Output levels Output low <0.8 V @ I = -4 mA
Output high >2.4 V @ I = 4 mA
Isolation None
Pulse Width 100 +/- 10 ms (high)
Delay from R-wave
peak to start of pulse
20 ms maximum per AAMI EC13
Minimum required R-
wave amplitude
0.5 V
Wireless Network
Device Interface
Signals RD+/-, TD+/-: IEEE 802.3 10Base-T, PWR, GND
12.5 V ±20%, 3.5 W continuous
Short Range Radio
Interface
1
Type Internal SRR interface
Technology IEEE 802.15.4
Frequency Band 2.4 GHz ISM (2.400 - 2.483 GHz)
Modulation Technique DSSS (O -QPSK)
Effective radiated
power
max. 0 dBm (1 mW)
M8023A External Power Supply (for M3002A) Interface Specifications
Measurement Link
(MSL)
Connectors Male ODU (Proprietary)
Power 48 V output
Power Sync. RS-422 compliant output 78.125 kHz (typical)
LAN signals IEEE 802.3 10-Base-T compliant
Serial signals RS-422 compliant output 78.125 kHz (typical)
Local signals Not connected
Integrated PC (iPC) Interface Specifications
Ethernet LAN (external) Connector RJ-45
LAN Signals IEE 802.3 1000-Base-T compliant
Reinforced insulation IEC60601-1 A-k compliant
USB 6 external ports (5 rear, 1 right side)
Type A connectors
USB 2.0 supporting high speed mode
Audio microphone input stereo 3.5 mm audio jack
headphone output stereo 3.5 mm audio jack
DVI Video with DVI-I connector DVI supports resolutions up to
1920x1200
VGA supports resolutions up to
2048x1536
X2 (M3002A) Interface Specifications