17 Monitoring EEG
250
Impedance Indicators
About Compressed Spectral Arrays (CSA)
The continuous EEG signal is sampled periodically and this value is stored in a frame. Each frame is
processed using Fast Fourier Transformation (FFT) to provide a frequency spectrum displayed as a
compressed spectral array (CSA).
The CSA window provides an overview of the patient's EEG values over time. It may be configured to
look slightly different on your monitor.
To view the CSA window permanently embedded on a Screen, select a Screen configured to
display the CSA window.
To view the CSA window over the current Screen, select the
EEG CSA SmartKey, if configured, or
select
Setup EEG or Main Setup then Show CSA.
Electrode/Skin
Impedance
Symbol Color Displayed
Impedance
Value
Action
Electrode not connected red no value connect electrode
Noisy signal gray 60 k (fixed) check electrode-to-skin
connections
Electrode connected,
impedance above limit
yellow measured value
(e.g. 15 k)
check limit, check
electrode-to-skin contact
Electrode connected,
impedance at or below limit
green measured value
(e.g. 3 k)
no action necessary