EasyManua.ls Logo

Philips IntelliVue MX800 - Making the Pressure Calibration; WARNING; 1 Zero the transducer.; 2 Connect the syringe and manometer.

Philips IntelliVue MX800
460 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11 Monitoring Invasive Pressure
188
approximately 25cm of tubing.
Making the Pressure Calibration
WARNING
Never perform the invasive pressure calibration while a patient is being monitored.
1 Zero the transducer.
2 Connect the syringe and manometer.
a. Attach the tubing to the manometer.
b. Connect the 3-way stopcock to the stopcock that is not connected to the patient catheter
when you measure a patient.
c. Attach the syringe to one port and the manometer tubing to the other port.
d. Open the port to the manometer.
1 Tubing to manometer
2 Syringe with heparinised solution
3 To pressure connector on monitor
4 Patient connection stoppered
5 Off
3 Move the syringe barrel in and raise the mercury to 200 mmHg (30 kPa). 200 mmHg is the
recommended calibration pressure.
4 In the Setup <Press Label> menu, select Cal. Press.
5 Select the calibration pressure from the list, for example 200 mmHg.
6 Select Confirm to recalculate the calibration factor using the applied pressure.
7 When the monitor displays <Press Label> mercury calibr. done at <date and time>, remove
the manometer tubing, syringe and extra stopcock. We recommend you replace the transducer
dome and tubing with sterile ones.
8 Label the transducer with the calibration factor shown in the Cal. Factor field in the pressure's
setup menu.

Table of Contents

Related product manuals