4 Managing Patients
99
– wait until the transfer has completed.
4 Connect the X2/MP5 to the new host monitor. The monitor detects a patient mismatch and a
window will open showing your patient's data and asking
Complete Transfer Of This Patient?.
5 Select Yes to complete the transfer. This re-admits the patient from the transfer list to the new
monitor. This will upload the patient demographics, and, if configured, the measurement settings
and trend data stored in the X2/MP5 to the receiving monitor.
6 Verify that the settings for patient category and paced mode are correct.
If you accidentally transfer a patient, use
Re-Admit to restore this patient's data to the Information
Center. If you are not connected to the network, select
Clear Transfer to leave transfer mode. The
patient data remains in the monitor.
Transferring a Patient with an X1 (no Information Center)
Scenario: A patient is moved to another monitoring location using a transport monitor and re-admitted
at the new monitor.
1 Disconnect the X1 from the original monitor and silence the resulting UNPLUGGED INOP.
If you are sure that the patient will not be returning to this monitor, you should perform a
discharge or end-case. This prevents data from the next patient accidentally being mixed up with
your current patient's data.
2 Connect the X1 to the transport monitor and move the patient.
3 At the new monitoring location, remove the X1 from the transport monitor and connect it to the
new monitor.
4 If prompted, re-admit the patient to the new monitor: in the Select Patient window, select the
patient in the X1 to retain the data in the X1. This will upload the patient demographics, and, if
configured, the measurement settings and trend data stored in the X1 to the monitor. Verify that
the settings for patient category and paced mode are correct.
Transferring a Patient with an X2 or MP5 (no Information Center)
WARNING
Measurements from a MMS extension connected to an X2 are not available when the X2 is running on
battery power. They are only available when the X2 is powered from AC mains, either when connected
to a host monitor or the external power supply (M8023A).
Scenario: A patient is moved to another monitoring location using the X2 or MP5 and re-admitted at
the new monitor.
1 Disconnect the X2/MP5 from the original host monitor and silence the resulting UNPLUGGED
INOP.
If you are sure that the patient will not be returning to this monitor, you should perform a
discharge or end-case. This prevents data from the next patient accidentally being mixed up with
your current patient's data.
2 Move the patient using the X2/MP5 as the transport monitor.
3 At the new monitoring location, connect the X2/MP5 to the new host monitor.
4 If prompted, re-admit the patient to the new monitor: in the Select Patient window of the new
monitor, select the patient in the X2/MP5 to retain the data in the X2/MP5. This will upload the