36 Specifications
424
Cardiac Output / Continuous Cardiac Output
tcGas
Complies with IEC 60601-2-23:1999/EN60601-2-23:2000.
C.O./CCO Performance Specifications
C.O. (right heart)
Range 0.1 to 20.0 l/min
Accuracy Instrument Specification (electrical): ±3% or 0.1 l/min
System Specification: ±5%.or 0.2 l/min
Repeatability Instrument Specification (electrical): ±2% or 0.1 l/min
System Specification: ±3%.or 0.1 l/min
C.O. (transpulmonary
)
Range 0.1 to 25.0 l/min
Accuracy Instrument Specification (electrical): ±4% or 0.15 l/min
System Specification: ±5% or 0.2 l/min
Repeatability Instrument Specification (electrical): ±2% or 0.1 l/min
System Specification: ±3% or 0.1 l/min
EVLW Range 10 to 5000 ml
Standard Deviation 10% or 1 ml/kg
ITBV Range 50 to 6000 ml
Accuracy ±10% or 30 ml
Repeatability ±5% or 20 ml
CCO Range 0.1 to 25.0 l/min
Standard Deviation 10% or 0.3 l/min
Display Update 2 seconds nominal
Blood Temperature Range 17.0 to 43°C (62.6 to 109,5°F)
Injectate Temperature Range -1 to 30°C (30.2 to 86.0°F)
C.O./CCO Alarm
Specifications
Range Adjustment Delay
TBlood 17 to 43°C Steps of 0.5°C (17 to 35°C)
Steps of 0.1°C (35 to 43°C)
Steps of 1°F (63 to 95°F)
Steps of 0.2°F (95 to 109°F)
10 seconds after the value
exceeds the set limit range
CCO 0.1 to 25.0 l/min 0.1 l/min (0.1 to 10.0 l/min)
0.5 l/min (10.0 to 25.0 l/min)
10 seconds after the value
exceeds the set limit range
tcGas Performance Specifications
tcpO2 Range 0 to 750 mmHg (0 to 100 kPa)
Resolution 1 mmHg (0.1 kPa)
Overall Accuracy
(incl. transducer)
0 to 20.9% O
2
better than +5 mmHg (+0.7 kPa)
Above 20.9% O
2
better than +10% of reading
tcpO2 Test Signal 60 mmHg (8.0 kPa)