Philips IntelliVue MX800 - Use Models With Telemetry; 1 The monitor is paired with a telemetry transceiver - indirect connection; 2 An MP5 or X2 is declared as a telemetry device and paired with an MX800 monitor - direct or indirect connection; 3 A telemetry transceiver directly connected (SRR or for MP5 cable) to an MP5 or X2, that is connected (Companion Mode) to an MX800 monitor - direct connection.

Philips IntelliVue MX800
460 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19 Assigning a Telemetry Device and a Monitor to One Patient
268
Use Models With Telemetry
The standard use model combining a monitor and a telemetry device involves pairing the two devices
so that the data measured by the telemetry device appears on the monitor screen - and at the
Information Center in the same patient sector as the monitor data. The following variations are
possible:
1 The monitor is paired with a telemetry transceiver - indirect connection
telemetry data appear on the monitor
2 An MP5 or X2 is declared as a telemetry device and paired with an MX800 monitor - direct or
indirect connection
MP5/X2 measurement data appear on the monitor it is paired with
a telemetry transceiver cannot be used with the MP5/X2
3 A telemetry transceiver directly connected (SRR or for MP5 cable) to an MP5 or X2, that is
connected (Companion Mode) to an MX800 monitor - direct connection.
telemetry data appear on the host monitor
telemetry transceiver is paired with the host monitor but is assigned to the MP5 or X2
the MP5 or X2 must have no equipment label
The different variations require different configuration settings; refer to the Configuration Guide for