19 Assigning a Telemetry Device and a Monitor to One Patient
268
Use Models With Telemetry
The standard use model combining a monitor and a telemetry device involves pairing the two devices
so that the data measured by the telemetry device appears on the monitor screen - and at the
Information Center in the same patient sector as the monitor data. The following variations are
possible:
1 The monitor is paired with a telemetry transceiver - indirect connection
– telemetry data appear on the monitor
2 An MP5 or X2 is declared as a telemetry device and paired with an MX800 monitor - direct or
indirect connection
– MP5/X2 measurement data appear on the monitor it is paired with
– a telemetry transceiver cannot be used with the MP5/X2
3 A telemetry transceiver directly connected (SRR or for MP5 cable) to an MP5 or X2, that is
connected (Companion Mode) to an MX800 monitor - direct connection.
– telemetry data appear on the host monitor
– telemetry transceiver is paired with the host monitor but is assigned to the MP5 or X2
– the MP5 or X2 must have no equipment label
The different variations require different configuration settings; refer to the Configuration Guide for
details.