36 Specifications
404
Interface Specifications
MX800 Interface Specifications
Network Standard 100-Base-TX (IEEE 802.3 Clause 25)
Connector RJ45 (8 pin)
Isolation basic insulation (reference voltage: 250 V; test voltage: 1500 V)
MIB/RS232 Standard IEEE 1073-3.2-2000
Connectors RJ45 (8 pin)
Mode Software-controllable
BCC (RxD/TxD cross over) or
DCC (RxD/TxD straight through)
Power 5 V ±5 %, 100 mA (max.)
Isolation basic insulation (reference voltage: 250 V; test voltage: 1500 V)
USB Interface
(4 ports)
Standard USB 2.0 full-speed (embedded host)
Connector USB series "Standard A" receptacle
Power Low power port 4.4V min; max. load for all ports together
500 mA
Isolation none
RS232 (Standard) Connector RJ45 (8-pin)
Power none
Isolation basic insulation (reference voltage: 250 V; test voltage: 1500 V)
RS232 (Independent
display interface
option)
Connector RJ45 (8-pin)
Power none
Isolation none
Basic Nurse Call
Relay
Connector modular Jack 6P6C, active open and closed contact
Contact <=100 mA, <=24 V DC
Isolation basic insulation (reference voltage: 250 V; test voltage: 1500 V)
Delay <[Configured Latency +0.5] sec
IntelliVue Instrument
Telemetry Wireless
Network (USA only)
Type Internal WMTS Adapter
Technology compatible with Philips IntelliVue Telemetry System (ITS),
cellular infrastructure
Frequency Band WMTS, 1395-1400 MHz and 1427-1432 MHz
IntelliVue Instrument
Telemetry Wireless
Network (except USA)
Type Internal ISM Adapter
Technology compatible with Philips IntelliVue Telemetry System (ITS),
cellular infrastructure
Frequency Band 2.4 GHz ISM
IntelliVue 802.11
Bedside Adapter
(Wireless Network
Adapter)
Type Internal Wireless Adapter
Technology IEEE 802.11a/b/g
Frequency Band 2.4 GHz and 5 GHz ISM Band