37 Default Settings Appendix
441
ECG, Arrhythmia, ST and QT Default Settings
Yel.Al. Interval 20 sec
Alarm Low 42
Red Alarm Volume AlarmVol. +2
Yell.AlarmVolume AlarmVol. +0
Inop Volume AlarmVol. +0
AutoIncrease Vol 2 Steps
IncreaseVolDelay 20 sec
Keep Blinking No
Relay1 Sensitiv. R&Y&C
Relay2 Sensitiv. Red&Yellow
Relay3 Sensitiv. Red
CyanRelayLatency 5sec
Yel.RelayLatency 2sec
Alarm Text Standard
NoCentrMonMinVol On4
Alarm Settings Factory Default H10/H20/H40
H30 (deviations from H10/
H20/H40)
ECG Settings
Factory defaults H10/H20/H40 H30 (deviations from H10/H20/H40)
Adult Pedi Neo Adult Pedi Neo
High Limit 120 bpm 160 bpm 200 bpm
Low Limit 50 bpm 75 bpm 100 bpm
ECG/Arrhy Alarms On On On
Alarm Source Auto Auto Auto Auto
ECG OnOnOn
QRS Volume 111
Primary Lead II II II
Secondary Lead V V V
Analysis Mode Multi Lead Multi Lead Multi Lead
Lead Placement Standard Standard Standard
Mod.LeadPlacment Off Off Off
Filter Monitor Monitor Monitor Filter
AutoFilter Off Off Off
Default ECG Size Size X1 Size X1 Size X1
Color Green Green Green
Asystole Thresh. 4.0 sec 4.0 sec 3.0 sec
Δ ExtrTachy 20 bpm 20 bpm 20 bpm
Tachy Clamp 200 bpm 220 bpm 240 bpm