12 Monitoring Cardiac Output
205
C.O./CCO Curve Alert Messages
After each measurement trial, the monitor analyzes the thermodilution curve. If the curve appears
abnormal, a curve alert message appears in the procedure window. A question mark symbol ("
?") may
appear next to the cardiac output numeric if any of these messages appear. The curve alert messages
are not error messages and do not necessarily mean that the measured values are invalid.
Patient
Weight
Cold Injectate Room Temperature Injectate
ETVI < 10 ETVI 10 ETVI < 10 ETVI 10
< 3 kg 2 ml 2 ml 3 ml Use cold injectate
< 10 kg 2 ml 3 ml 3 ml
< 25 kg 3 ml 5 ml 5 ml
< 50 kg 5 ml 10 ml 10 ml
< 100 kg 10 ml 15 ml 15 ml
100 kg 15 ml 20 ml 20 ml