14
223
14Monitoring Airway Flow,
Volume and Pressure
The M1014A Spirometry Module is designed to be used in combination with Philips-branded airway
flow sensors and combined CO
2
/airway flow sensors. It produces a real time wave for flow, volume
and pressure of respiratory gases together with numerics for analysis of ventilatory mechanics.
The measurement provides:
Airway flow, airway volume and airway pressure waveform.
•Numerics for:
respiratory rate
–tidal volume
minute volume
PEEP (positive end-expiratory pressure)
PIP (peak inspiration pressure)
dynamic compliance
airway resistance
Numerics for: (only visible in
SpirometryValues Window, no alarming, no trending)
–peak inspiratory flow
–peak expiratory flow
negative inspiratory pressure
mean airway pressure
plateau pressure numeric
–I:E ratio
The measurement also provides pressure-volume loops, flow-volume loops and pressure-flow loops.
Please see the “Respiratory Loops” on page 355 chapter for details on loops.
When using combined CO
2
/Flow sensors you can also perform mainstream capnography with
intubated adult, pediatric and neonatal patients. See the Measuring CO
2
chapter for details on the CO
2
measurement and the Accessories chapter for a list of compatible accessories.