36 Specifications
417
Invasive Pressure Alarm Specifications
Temp
Complies with EN 12470-4:2000. Specified without transducer.
CO2
The CO
2
measurement in M3014A and M3015A complies with EN ISO 21647:2004 + Cor.1:2005
(except alarm system; alarm system complies with IEC 60601-2-49:2001).
Invasive Pressure Alarm
Specifications
Range Adjustment Delay
Pressure -40 to 360 mmHg
(-5.0 to 48 kPa)
-40 to 50 mmHg
2 mmHg (0.5 kPa)
>50 mmHg
5 mmHg (1 kPa)
max.
12 seconds
Extreme High Difference to high limit 0 to
25 mmHg
5 mmHg steps (0.5 kPa)
Clamping at -40 to 360 mmHg 5 mmHg steps (1.0 kPa)
Extreme Low Difference to low limit 0 to
25 mmHg
5 mmHg steps (0.5 kPa)
Clamping at -40 to 360 mmHg 5 mmHg steps (1.0 kPa)
Pulse 25 to 300 bpm Adult:
1 bpm steps (25 to 40 bpm)
5 bpm steps (40 to 300 bpm)
Pedi/Neo:
1 bpm steps (25 to 50 bpm)
5 bpm steps (50 to 300 bpm)
Tachycardia Difference to high limit 0 to 50 bpm 5 bpm steps max.
14 seconds
Clamping at 150 to 300 bpm 5 bpm steps
Bradycardia Difference to low limit 0 to 50 bpm 5 bpm steps max.
14 seconds
Clamping at 25 to 100 bpm 5 bpm steps
Temp Performance Specifications
Temp Range -1 to 45°C (30 to 113°F)
Resolution 0.1°C (0.2°F)
Accuracy ±0.1°C (±0.2°F)
Average Time Constant
Less than 10 seconds
Temp Alarm Specifications Range Adjustment
Temp High/Low Alarms
-1 to 45°C (30 to 113°F) -1 to 35°C (30 to 95°F), 0.5°C (1.0°F) steps
35 to 45°C (95 to 113°F), 0.1°C (0.2°F) steps