36 Specifications
416
Invasive Pressure Performance Specifications
Measurement Range -40 to 360 mmHg
Pulse Rate Range 25 to 350 bpm
Accuracy ±1% Full Range
Resolution 1 bpm
Input Sensitivity Sensitivity: 5 µV/V/mmHg (37.5 µV/V/kPa)
Adjustment range: ±10%
Transducer Load Impedance: 200 to 2000 Ω (resistive)
Output Impedance: ≤3000 Ω (resistive)
Frequency Response dc to 12.5 Hz or 40 Hz
Zero Adjustment Range ±200 mmHg (±26 kPa)
Accuracy ±1 mmHg (±0.1 kPa)
Drift Less than 0.1 mmHg/°C (0.013 kPa/°C)
Gain Accuracy Accuracy ±1%
Drift Less than 0.05%/°C
Non linearity and
Hysteresis
Error of ≤0.4 %FS (@CAL 200 mmHg)
Overall Accuracy (including
transducer)
±4% of reading or ±4 mmHg (±0.5 kPa), whichever is greater
Volume displacement of CPJ840J6 0.1 mm
3
/100 mmHg
Analog output Specifications
Analog Output
available only with
M1006B #C01 (@ CAL
200 mmHg)
Range -0.4 V to 3.6 V
Level 1 V / 100 mmHg
Accuracy ±3% full scale
Offset ±30 mV
Resolution 8 Bit (@ 5 V range)
Signal delay 20 ms