2Alarms
55
1 Measurement labels, with alarms off symbol where appropriate
2 Current alarm limits
3 Graphic view of current yellow and red alarm limits and currently monitored measurement value.
Off indicates the measurement is switched off
•Select
Show ST Limits to expand the list of ST leads and view the currently set alarm limits.
Selecting
Hide ST Limits hides the list again.
You can use the pop-up keys that open with the
Alarm Limits window to perform common tasks:
All Al. On/All Al. Off
All Lim. Narrow/All Lim. Wide to set narrow or wide alarm AutoLimits for all measurements.
Print Limits/Record Limits to print a list of all current alarm limit settings on a connected
printer or recorder.
These pop-up keys are not available in the window for changing individual alarm limits which you
access by selecting the measurement label in the
Alarm Limits window.
Changing Alarm Limits
To change individual measurement alarm limits using the measurement's Setup Menu,
1 In the measurement's setup menu, select the alarm limit you want to change. This calls up a list of
available values for the alarm limit.
2 Select a value from the list to adjust the alarm limit.
Alternatively, you can use the keys in the measurement alarm limits window, which you access by
selecting the measurement label in the
Alarm Limits window.