37 Default Settings Appendix
449
Spirometry Default Settings
tcGas Default Settings
Spirometry Settings Factory Adult Factory Pedi Factory Neo
Apnea Time 20 20 20
MVexp Low 4.0 l/min 2.5 l/min 0.4 l/min
MVexp High 8.0 l/min 4.0 l/min 0.8 l/min
PIP High 40 25 20
PEEP Low 000
PEEP High 25 25 25
High Limit (RRspir)8830
Low Limit (RRspir) 303060
Scale (AWF) ±150 l/min ±100 l/min ±20 l/min
Scale (AWV) ±800 ml ±200 ml ±50 ml
Scale (AWP) ±40 cmH
2
O ±40 cmH
2
O ±20 cmH
2
O
tcGas Settings Factory Adult Factory Pedi Factory Neo
tcpO High Limit 80 mmHg 80 mmHg 80 mmHg
tcpO Low Limit 50 mmHg 50 mmHg 50 mmHg
tcpO Alarms On On On
tcpCO HighLimit 50 mmHg 50 mmHg 50 mmHg
tcpCO Low Limit 30 mmHg 30 mmHg 30 mmHg
tcpCO Alarms On On On
Site Timer 4.0 hours 4.0 hours 4.0 hours
Disable Timer Not Allowed Not Allowed Not Allowed
Heat Switch Off (i.e., after Site Timer
elapsed)
No No No
Transducer Temp. 43.0°C 43.0°C 43.0°C
CO Correction (Severinghaus) On On On
MetabolismFactor 8 mmHg 8 mmHg 8 mmHg
tcGas Unit mmHg mmHg mmHg
Temperature Unit °C °C °C
tcpO Color Blue Blue Blue
tcpCO Color Green Green Green