3 Patient Alarms and INOPs
91
Cableless Measurement Device INOPs
INOP Message, Indication What to do
cl NBP Batt Empty The remaining battery time of the NBP Pod is below 30 minutes. Change the battery.
cl NBP Batt Incomp The battery in use with the NBP Pod is incompatible. Replace it with one approved for use with the
NBP Pod.
cl NBP Batt Low The remaining battery time of the NBP Pod is below 2 hours.
cl NBP Batt Malf There is a malfunction in the NBP Pod's battery system. Contact your service personnel.
cl NBP Batt Temp The temperature of the battery in the NBP Pod is critically high. Check that the Pod is not covered or
exposed to a heat source. If the INOP persists, remove the Pod from the patient and contact your
service personnel.
cl NBP Check Batt The battery in the NBP Pod is nearing the end of its useful life. Only 50 charge/discharge cycles
remain. Contact your service personnel to replace the battery.
cl NBP Disconnect The NBP Pod has lost the connection to the monitor.
cl NBP No Cradle The NBP Pod is not in its cradle.
cl NBP Remove The temperature of the battery in the NBP Pod is too high. Remove the Cableless Measurement
Device from the patient and contact service personnel.
cl NBP Serv Batt The battery in the NBP Pod has reached the end of its useful life. It can no longer be charged. Contact
your service personnel to replace the battery
cl SpO Batt Empty The remaining battery time of the SpO
2
Pod is below 30 minutes. Change the battery.
cl SpO Batt Incmp The battery in use with the SpO
2
Pod is incompatible. Replace it with one approved for use with the
SpO
2
Pod.
cl SpO Batt Low The remaining battery time of the SpO
2
Pod is below 2 hours.
cl SpO Batt Malf There is a malfunction in the SpO
2
Pod's battery system. Contact your service personnel.
cl SpO Batt Temp The temperature of the battery in the SpO
2
Pod is critically high. Check that the Pod is not covered or
exposed to a heat source. If the INOP persists, remove the Pod from the patient and contact your
service personnel.
cl SpO Check Batt The battery in the SpO
2
Pod is nearing the end of its useful life. Only 50 charge/discharge cycles
remain. Contact your service personnel to replace the battery.
cl SpO Disconnect The SpO
2
Pod has lost the connection to the monitor.
cl SpO No Cradle The SpO
2
Pod is not in its cradle.
cl SpO Remove The temperature of the battery in the SpO
2
Pod is too high. Remove the SpO
2
Pod from the patient
and contact service personnel.
cl SpO Serv Batt The battery in the SpO
2
Pod has reached the end of its useful life. It can no longer be charged. Contact
your service personnel to replace the battery