3 Patient Alarms and INOPs
89
IntelliBridge INOPs
Telemetry INOPs
INOP Message, Indication What to do
DEVICE CHECK CONF.
INOP tone
Device identification completed, but communication could not be established due to error.
IntelliBridge INOP abbreviations may differ slightly depending on the device category.
DEVICE CHECK SETUP
INOP tone
Device identification completed, but communication could not be established due to timeout.
IntelliBridge INOP abbreviations may differ slightly depending on the device category.
DEVICE DEMO DATA
INOP tone
The device connected to the IntelliBridge module reports demo data but the monitor is not in DEMO
mode.
DEVICE REAL DATA
INOP tone
The monitor is in DEMO mode but the device connected to the IntelliBridge module reports data that
are not flagged as demo data.
DEVICE UNSUPPORTED
INOP tone
Device identification completed, but no appropriate device driver installed.
IntelliBridge INOP abbreviations may differ slightly depending on the device category.
<EC10 / EC40> EQUIP
MALF
INOP tone
Malfunction in the IntelliBridge module. If this message appears repeatedly, the module must be
replaced. Contact your service personnel.
EC10 INOP
!! EC10 INOP
!!! EC10 INOP
at Information Center
A technical alarm is present on the IntelliBridge EC10 module.
On the monitor the indication is a red (!!!), yellow (!!) or cyan alarm lamp (as appropriate) and an INOP
text that is defined by the IntelliBridge EC10 device driver.
<External Device>
UNPLUGGED
INOP tone
The IntelliBridge module has been unplugged from the rack, or the whole rack has been disconnected.
Silencing this INOP switches off the measurement.
IntelliBridge INOP abbreviations may differ slightly depending on the device category.
NO DEVICE DATA Communication with connected device has been lost.
TEXT UPLOAD FAILED
INOP tone
Incoming text from the IntelliBridge modules exceeds the maximum limit. Try unplugging one of the
IntelliBridge modules. If the INOP occurs repeatedly contact your service personnel; a software
upgrade may be necessary.
INOP Message, Indication What to do
!! REPLACE BATT. T
!!!REPLACE BATT. T
Severe yellow/red INOP tone
During this INOP, alarms cannot
be paused or switched off.
The battery in the telemetry device is almost empty and must be replaced.
!! TELE INOP
!!! TELE INOP
Severe yellow/red INOP tone
Check for further details at the Information Center or in the Telemetry Data window on the monitor.
BATTERY LOW T The battery in the Telemetry device is low and must be replaced soon.
Check ECG Settings
INOP tone
Synchronization of ECG settings between the monitor and Information Center has failed. Check that
the ECG settings in use are appropriate.
!!CHECK ECG SOURCE Both the telemetry device and the monitor have valid ECG signals
Chk SpO₂T Settings
INOP tone
Synchronization of SpO
2
T settings between the monitor and Information Center has failed. Check
that the SpO
2
T settings in use are appropriate.