24 ProtocolWatch
309
CAUTION
It is not possible to transfer SSC Sepsis Protocol data from a monitor with release F.0 software to
another monitor with release G.0 software or above, and vice versa.
Suspending the Protocol for 24 Hours
To suspend the protocol for 24 hours:
1 Select Main Setup then ProtocolWatch followed by Suspend for 24 hours.
2 Select Confirm.
The protocol will resume activity after 24 hours with Severe Sepsis Screening.
Severe Sepsis Screening
As soon as an adult patient is connected to the monitor, and heart rate or pulse is being measured, the
protocol begins to observe heart rate, temperature and respiration rate values in the background. If
temperature or respiration rate is not being measured continuously, a pop-up window appears at 4-
hour intervals to ask for manually measured values. These values are all compared to the screening
criteria: HR above 90, Temp above 38.3°C (100.9°F) or below 36.0°C (96.8°F), RR above 20.
If any one of the values fulfills the screening criteria, the first
Severe Sepsis Screening window
appears:
The measured value which has fulfilled the criteria is shown highlighted (in the example shown above
it is respiration). There is a check box for the infection signs and symptoms listed,
• if the monitor has measured values which meet the screening criteria, and
• for all signs and symptoms which cannot be measured by the monitor.
Check the box if you can confirm that this sign or symptom is present and new. If the patient is
mechanically ventilated, the
Mechanically ventilated box must be checked. If you have manually
measured temperature or respiration rate, you can enter the values using the
Enter RR and Enter Temp
keys.