14 Monitoring Airway Flow, Volume and Pressure
228
NOTE
If you unintentionally switch the gas compensation to the incorrect setting, the accuracy of the
measured flow and volume values may be significantly reduced. The tables below show examples of
incorrect gas compensation settings and the resulting inaccuracies:
Incorrect Balance Gas Setting (N
2
O instead of N
2
)
Incorrect Agent Percentage Setting
Setting up Spirometry
The following settings can be accessed through the Setup Spirometry menu.
Optimizing Scale Settings
In the Setup Spirometry menu, select Optimize Scales. The scale settings for all three spirometry
waves (AWF, AWP, AWV) will be optimized.
Spirometry Values Window
You can view all spirometry measurement data in the spirometry data window. In the Setup
Spirometry
menu, select Show All Values.
Alarms and Alarm Limits
Alarms can be derived from PEEP, PIP, MV and RRspir. In the Setup Spirometry menu you can set
the alarm limits. Please refer to the Alarms chapter of this manual for details on alarm settings.
Changing the Apnea Alarm Delay
This determines the time limit after which the monitor alarms if the patient stops breathing.
1 In the Setup Spirometry menu, select RRspir.
2 In Setup RRspir menu, select Apnea Time.
3 Choose the apnea alarm delay time.
N
2
O
2
N
2
O Agent Difference between
measured and actual
values
Actual Gas Composition
60 40
Gas Compensation Setting
40 60 15%
N
2
O
2
N
2
O Agent Difference between
measured and actual
values
Actual Gas Composition
55 40 5
Gas Compensation Setting
60 40 0 15%