3 Patient Alarms and INOPs
88
VueLink INOPs
SPIRO NO BREATH No breath was detected for more than 25 seconds. Breath derived numerics are not available.
SPIRO NO SENSOR No sensor detected. Make sure the correct sensor is attached to the breathing circuit.
SPIRO PATIENT CAT. Mismatch of patient size configured in the host monitor and sensor type plugged into the module.
Check the instructions on selecting the correct sensor in the chapter on “Spirometry” on page 422.
SPIRO PURGE FAILED The purge operation could not be completed successfully. Check for kinked sensor tubings, hard
occlusions and make sure that the pump is running and all valves are switching.
SPIRO PURGING A purge operation is in progress - no data update on the screen. Wait until purge is complete.
SPIRO UNKN. SENSOR An unknown sensor ID code was detected. Use only the sensors listed in the “Accessories” on
page 369 chapter.
SPIRO UPGRADE The module is running a firmware upgrade. Wait until upgrade is completed before resuming
monitoring.
INOP Message, Indication What to do
INOP Message, Indication What to do
VueLink ALARM
at Information Center
A technical alarm is present on the VueLink module. The INOP text on the monitor is defined by the
VueLink device driver.
<VueLink Option> CHECK
SETUP
INOP tone
No information was received from the external device. The device may be switched off or
disconnected.
VueLink INOP abbreviations may differ slightly depending on the device category.
<VueLink Option> CHK
CABLE
INOP tone
No cable or the wrong cable connected to the VueLink module, or incorrect device selected. Silencing
this INOP switches the measurement off.
VueLink INOP abbreviations may differ slightly depending on the device category.
<VueLink Option> CHK
CONF.
INOP tone
The wrong external device has been selected on the VueLink module, or the external device has not
been correctly setup, or the wrong cable has been used to connect the device to the VueLink module.
VueLink INOP abbreviations may differ slightly depending on the device category.
<VueLink Option> EQUIP
MALF
INOP tone
Malfunction in the VueLink module. If this message appears repeatedly, the module must be replaced.
Contact your service personnel.
VueLink INOP abbreviations may differ slightly depending on the device category.
<VueLink Option> NO
CONFIG
INOP tone
The VueLink module has not been configured during installation. The installation process should be
completed by either your biomedical engineering department or the Philips service engineer.
VueLink INOP abbreviations may differ slightly depending on the device category.
<VueLink Option>
UNPLUGGED
INOP tone
The VueLink module has been unplugged from the rack, or the whole rack has been disconnected.
The measurement automatically disappears from the display. Silencing this INOP switches off the
measurement.
VueLink INOP abbreviations may differ slightly depending on the device category.