11 Monitoring Invasive Pressure
187
Select and set the High Limit and Low Limit for the pressure(s) you have selected.
Extreme Alarm Limits
The extreme pressure alarms, Extreme High and Extreme Low, can be made available for your
monitor in Configuration Mode and are additional to the standard High and Low limit alarms. They
are generated by the active pressure alarm source, and are setup in Configuration Mode by adding a set
value (the Δ value) to the high and low alarm limits. This value can be set for each pressure label
individually.
1 Extreme Low Limit
2 Low Limit
3 High Limit
4 Extreme High Limit
5 Δ Extreme Low
6 Δ Extreme High
You need to know which values have been configured for your monitor. Changing the high and low
alarm limits automatically changes the extreme alarm limits within the allowed range.
• To see the extreme pressure alarms set for your monitor, in the
Setup <Press Label> menu, see
the menu items
Δ Extreme High and Δ Extreme Low.
The extreme pressure alarms are high priority, red alarms, marked
*** in the alarm message.
Calibrating Reusable Transducer CPJ840J6
Depending on your monitor's configuration, you may be able to perform a calibration in monitoring
mode. Perform a mercury calibration when you use a new transducer, and at regular intervals
according to your hospital policy. You require:
• standard sphygmomanometer.
• sterile 10cc syringe with heparinised solution.
•3-way stopcock.
Sys & Dia systolic and diastolic in parallel
Dia & Mean diastolic and mean in parallel
Sys & Mean systolic and mean in parallel
Sys&Dia&Mean all three pressures in parallel
Menu option Pressure value monitored