EasyManua.ls Logo

Philips IntelliVue MX800 - Preparing to Measure Mainstream CO2; 1 Attach the transducer connector to the CO2 connector on the M3016 A extension.; 2 Wait 20 minutes, allowing the transducer to reach its operating temperature and a stable thermal condition.; 3 Perform an accuracy check and then, if necessary, calibrate the transducer.

Philips IntelliVue MX800
460 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13 Monitoring Carbon Dioxide
217
Preparing to Measure Mainstream CO2
1 Attach the transducer connector to the CO
2
connector on the M3016A extension.
2 Wait 20 minutes, allowing the transducer to reach its operating temperature and a stable thermal
condition.
3 Perform an accuracy check and then, if necessary, calibrate the transducer.
Checking Transducer Accuracy
WARNING
Check transducer accuracy at least once a week or if you doubt the CO
2
readings.
1 In Setup CO menu, select Cal. Mode to switch on calibration mode.
2 Look at the calibration value displayed in the Setup CO menu next to Start CAL1. Is it the same
as the value on the calstick? If not, calibrate the transducer now.
3 Place the transducer on the low cell of the calstick (labeled 0.0 mmHg or "ZERO"). The reading
on the screen should be zero within ±1 mmHg within one minute.
4 Place the transducer on the high cell of the calstick. The reading on the screen should be within ±1
mmHg of the value on the calstick within one minute.
5 If both readings are in range, you can leave calibration mode and begin monitoring. If either of the
readings is out of range, calibrate the transducer.
Calibrating the Transducer
1 Check that the windows on the calstick are clean and clear.
2 Place the transducer on one of the calstick cells and select Start CAL1.
3 Enter the calibration value printed on the calstick then press Confirm to start calibration.
4 When the message CO CAL1 calibration done -- Start CAL2 calibration appears, put the
transducer on the other cell and select
Start CAL2 then press Confirm.
5 When you see the message CO calibration completed - leave Cal. Mode, calibration is
complete.
6 Select Cal. Mode to switch calibration mode off. You cannot monitor in calibration mode.
Attaching and Removing the CO2 Transducer
1 Open the latch and place the transducer (B) onto the airway adapter (A). Place the airway adapter
in the patient's breathing circuit between the endotracheal tube and the Y-piece. You may see the

Table of Contents

Related product manuals