17Monitoring EEG
The Electroencephalograph (EEG) module monitors the patient's cerebral function by measuring the
electrical activity of the brain. It provides the monitor with two channels of realtime EEG waves, EEG
trend information in the form of Compressed Spectral Arrays (CSA), and up to eight of the following
numerics:
Spectral Edge Frequency - The SEF is the frequency below which a defined percentage of the Total
Power lies. The percentage is set in Configuration Mode.
Mean Dominant Frequency (MDF) - The MDF is the mean value of the frequency which
dominates the measured EEG.
Peak Power Frequency (PPF) - The PPF is the frequency with the highest measured amplitude.
Total Power (TP) - The TP numeric indicates the power in the measured frequency band.
Percentage of total power in each frequency band:
Alpha waves (8 to 13 Hz)
Beta waves (13 to 30 Hz)
Theta waves (4 to 8 Hz)
Delta waves (0.5 to 4 Hz).
EEG Monitoring Setup
1 Plug the trunk cable into the EEG module.
2 Prepare the patient's skin prior to placing the electrodes. Good electrode-to-skin contact is
important for a good EEG signal, as the skin is a poor conductor of electricity.
Shave hair from sites, if necessary.
Wash sites thoroughly with soap and water. We do not recommend using ether or pure
alcohol, because this dries the skin and increases the resistance.
Use a skin preparation paste to remove skin cells and oil before placing the electrodes.
3 Select the desired electrode montage in the Setup EEG menu or in the EEG Impedance / Montage
window.
4 Attach the reference electrode first.
5 Place the electrodes on the patient's head according to the selected montage. Use electrode gel if
you are not using pre-gelled electrodes. Remember to select a site where the signal will not be
interfered with by muscle artifacts.
6 Connect the electrode connector end to the trunk cable.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MX800 and is the answer not in the manual?

Philips IntelliVue MX800 Specifications

General IconGeneral
Resolution1280 x 1024 pixels
TouchscreenYes
Display TypeTFT LCD
Parameters MonitoredECG, SpO2, Temperature
Network ConnectivityEthernet, Wi-Fi
ConnectivityUSB, Ethernet
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C