36 Specifications
411
Respiration
SpO2
Unless otherwise specified, this information is valid for SpO
2
measured using the M3001A and
M3002A Multi-measurement modules and the M1020B measurement module. The SpO
2
Performance
Specifications in this section apply to devices with Philips SpO
2
technology. For SpO
2
Performance
Specifications valid for other SpO
2
technologies, refer to the instructions for use provided with these
devices.
Complies with ISO 9919:2005 / EN ISO 9919:2009 (except alarm system; alarm system complies with
IEC 60601-2-49:2001).
Measurement Validation: The SpO
2
accuracy has been validated in human studies against arterial
blood sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically
distributed, only about two-thirds of the measurements can be expected to fall within the specified
accuracy compared to CO-oximeter measurements.
Display Update Period: Typical: 2 seconds, Maximum: 30 seconds. Maximum with NBP INOP
suppression on: 60 seconds.
Respiration Performance Specifications
Respiration Rate Range Adult/pedi: 0 to 120 rpm
Neo: 0 to 170 rpm
Accuracy at 0 to 120 rpm ±1 rpm
at 120 to 170 rpm ±2 rpm
Resolution 1 rpm
Bandwidth 0.3 to 2.5 Hz (-6 dB)
Noise Less than 25 mΩ (rms) referred to the input
Respiration Alarm
Specifications
Range Adjustment Delay
High Adult/pedi: 10 to 100
rpm
Neo: 30 to 150 rpm
under 20 rpm: 1 rpm steps
over 20 rpm: 5 rpm steps
max. 14 seconds
Low Adult/pedi: 0 to 95 rpm
Neo: 0 to 145 rpm
under 20 rpm: 1 rpm steps
over 20 rpm: 5 rpm steps
for limits from 0 to 20 rpm:
max. 4 seconds
for limits above 20 rpm: max.
14 seconds
Apnea Alarm 10 to 40 seconds 5 second steps