3 Patient Alarms and INOPs
77
SpO2 INOPs
<Temp Label>
UNPLUGGED
INOP tone
A Temp measurement label has been deactivated, either by unplugging a module, or by deactivating
the label in the Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either replug the module or reactivate the measurement label in
the Measurement Selection window.
Tesoph INOPs See <Temp Label> INOPs
Tnaso INOPs See <Temp Label> INOPs
Trect INOPs See <Temp Label> INOPs
Tskin INOPs See <Temp Label> INOPs
Ttymp INOPs See <Temp Label> INOPs
Tven INOPs See <Temp Label> INOPs
Tvesic INOPs See <Temp Label> INOPs
INOP Message, Indication What to do
INOP Message, Indication What to do
<SpO Label>
DEACTIVATED
INOP tone
The SpO
2
measurement label in the measurement device has been deactivated by deactivating the label
in the Measurement Selection window. The measurement automatically disappears from the
display. To switch the measurement on again, reactivate the measurement label in the Measurement
Selection window.
<SpO Label> EQUIP MALF
Numeric is replaced by -?-
INOP tone
The MMS or module is faulty. Unplug and replug the MMS or module. If the INOP persists, contact
your service personnel.
<SpO Label> ERRATIC
Numeric is replaced by -?-
INOP tone
Check the sensor placement. Try another adapter cable and sensor. If the INOP persists, contact your
service personnel.
<SpO Label>
EXTD.UPDATE
Numeric is replaced by -?-
(questionable numeric)
The update period of displayed values is extended due to an NBP measurement on the same limb or
an excessively noisy signal.
<SpO Label>
INTERFERNCE
Numeric is replaced by -?-
INOP tone
There is too much interference, caused by a high level of ambient light and/or electrical interference.
Cover the sensor to minimize ambient light. If the INOP persists, make sure that the sensor cable is
not damaged or positioned too close to power cables.
<SpO Label> LOW PERF
Numeric is replaced by -?-
(questionable numeric)
Accuracy may be compromised due to very low perfusion. Stimulate circulation at sensor site. If INOP
persists, change the measurement site.
<SpO Label> NO SENSOR
Numeric is replaced by -?-
INOP tone
Make sure the SpO
2
sensor is connected. If the INOP persists, try another adapter cable and sensor. If
you silence this INOP, the measurement will be switched off.
<SpO Label> NOISY SIGN.
Numeric is replaced by -?-
INOP tone
Excessive patient movement or electrical interference is causing irregular pulse patterns. Try to reduce
patient movement or to relieve the cable strain on the sensor.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MX800 and is the answer not in the manual?

Philips IntelliVue MX800 Specifications

General IconGeneral
Resolution1280 x 1024 pixels
TouchscreenYes
Display TypeTFT LCD
Parameters MonitoredECG, SpO2, Temperature
Network ConnectivityEthernet, Wi-Fi
ConnectivityUSB, Ethernet
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C