37 Default Settings Appendix
445
Temperature Default Settings
Invasive Pressure Default Settings
Temp Settings
Factory defaults H10/H20/H40 H30 (deviations from H10/H20/H40)
Adult Pedi Neo Adult Pedi Neo
Low Limit 36 36 36 35 35 35
High Limit 39 39 39
Alarms On On On
Unit °C °C °C
Range 35...43 35...43 35...43
Color Green Green Green Light
Green
Light
Green
Light
Green
ABP, ART, Ao, BAP,
FAP, P, P1, P2, P3, P4,
UAP Settings
Factory defaults H10/H20/H40 H30 (deviations from H10/H20/H40)
Adult Pedi Neo Adult Pedi Neo
Alarms from Sys. Sys. Sys.
High Limit 160/90 (110) 120/70 (90) 90/60 (70) 180/90 (110)
Low Limit 90/50 (70) 70/40 (50) 55/20 (36) 70/50 (70)
Alarms On On On
Extreme Alarms Disabled Disabled Disabled
Δ Extreme High 15 10 5
Δ Extreme Low 15 10 5
Sys. High Clamp 190/100
(125)
140/80 (100) 105/75 (75)
Sys. Low Clamp 80/45 (65) 60/35 (45) 45/15 (30) 65/45 (65)
Scale 150 100 100
Mean only No No No
Filter 12 Hz 12 Hz 12 Hz
Mercury Cal Yes Yes Yes
Artifact Suppr. 60 sec 60 sec 60 sec
Unit mmHg mmHg mmHg
Color Red Red Red
Pulse settings (for ABP, ART, Ao, BAP, FAP, UAP,P)
Pulse (Press label) On On On
For all other Pulse default settings, refer to the Pulse Default Settings table.