28
343
28IntelliBridge EC10 Module
An IntelliBridge EC10 module transmits information from a connected external device to your
monitor. The IntelliBridge EC5 ID module is used to provide identification information from the
external device. Data imported from the external device, for example waveforms, measurement
numerics, settings and alarms, can be displayed on the monitor, passed on to an Information Center
and included in trends as appropriate. The external device may show more information than is
available on the monitor. Data from IntelliBridge EC10 modules cannot be included in monitor trends
when the label is a free-text label.
See the documentation supplied with the IntelliBridge module for a list of supported devices and
accessories, and for configuration information.
Connecting an External Device
1 Insert the module into the FMS or integrated module slot.
2 Check that you have the correct EC5 ID module for the external device - indicated on the EC5
label.
3 Connect the EC5 ID module to the external device.
4 Connect the EC5 to the EC10 module using the supplied cable.
5 Switch the external device on.
6 The connection status LED (4) will flash green until it has correctly identified the external device
and started communication. Check that the connection status LED then lights green continuously
indicating that communication has been established.
1 module name
2 setup indicator LED
3 setup/data viewing key
4 connection status LED
5 external device cable connector