12 Monitoring Cardiac Output
198
Setting Up the PiCCO C.O. Measurement
1 Set up the arterial line using the arterial catheter (transpulmonary catheter) and the transducer kit
from PULSION Medical Systems. It must be placed in one of the bigger systemic arteries, for
example, the femoral or the axillary artery. You must use the approved catheters and puncture
locations.
2 Set up the central venous line.
3 Connect the injectate temperature probe housing to the venous line.
4 Plug the C.O. interface cable into the C.O. module or measurement extension module and connect
the following devices to the C.O. interface cable:
– Injectate temperature probe
– Thermistor connector
– Remote start switch (if used).
Follow your hospital standards to avoid unintentional extraction of the C.O. catheter. Secure the
cable using the mounting clip shipped with each C.O. interface cable. You may also find it helpful
to loop the C.O. interface cable, tape the loop, and attach it to the undersheet of the patient's bed
using a safety pin.
1C.O.Module
2C.O.Interface Cable
3 PULSION Pressure Transducer
4 Press Module
5 Pressure Adapter Cable
6 Thermistor Connection Cable
7 Thermistor Connection
8 PULSION Arterial Catheter
9CVP Line
10 Injectate Temperature Probe
Housing
11 Injectate Syringe
12 Injectate Temperature Probe
13 Remote Switch