37 Default Settings Appendix
443
Respiration Default Settings
SpO2 Default Settings
Respiration Settings
Factory defaults H10/H20/H40 H30 (deviations from H10/H20/H40)
Adult Pedi Neo Adult Pedi Neo
High Limit 30 rpm 30 rpm 100 rpm
Low Limit 8 rpm 8 rpm 30 rpm
Apnea Time 20 sec 20 sec 20 sec
Alarms On On On
Resp On On On Off Off Off
Detection Auto
(Trigger
Mode)
Auto
(Trigger
Mode)
Auto
(Trigger
Mode)
Color Yellow Yellow Yellow White White White
SpO
2
Settings
Factory Defaults
Adult Pedi Neo
Mode Continuous Continuous Continuous
Repeat Time 15 min 15 min 15 min
Alarms On On On
QRS Volume 111
Tone Modulation Yes Yes Yes
Tone Mod. Type Enhanced Enhanced Enhanced
Perfusion OnOnOn
Average 10 sec 10 sec 10 sec
NBP Alarm Suppr. OnOnOn
Extd. Auto OnOff Disabled Disabled Disabled
Color Cyan (light blue) Cyan (light blue) Cyan (light blue)
Average In Mon. No No No
Signal Quality OnOnOn
SpO
2
Alarm Default Settings
Setting Adult Pedi Neo
Desat Limit 80 80 80
Low Limit 90 90 85
High Limit 100 100 95
Desat Delay 20 sec 20 sec 20 sec