3 Patient Alarms and INOPs
81
SO2 INOPs
CO₂ NO TUBING
Numeric is replaced by -?-
INOP tone
Either the sample line is disconnected, or an incorrect line is attached. Check the connection. If
necessary, connect another sample line (Use only the approved accessories).
If you silence this INOP, the measurement will be switched off.
CO₂ OVERRANGE
Numeric is replaced by -?-
INOP tone
The CO
2
value is higher than the measurement range. If you suspect a false high value, contact your
service personnel.
CO₂ PUMP OFF
Numeric is replaced by -?-.
The pump has been switched off for fifteen minutes. To switch it on again, select Pump On in the
Setup CO₂ menu.
CO₂ PURGING
Numeric is replaced by -?-
INOP tone
The Filterline is being purged to remove an occlusion in the line or airway adapter. If the occlusion is
removed, the INOP will disappear. If not, the INOP CO₂ OCCLUSION is displayed.
CO₂ SENS.WARMUP
Numeric is replaced by -?-
Microstream CO
2
: INOP tone
Mainstream CO
2
: no INOP tone
Wait until the sensor reaches operating temperature and the INOP disappears.
CO₂ UPDATE FW
Numeric is replaced by -?-
INOP tone
The software in the Measurement Extension does not match the software in the MMS. Contact your
service personnel.
CO₂ WAIT CAL2
Numeric is replaced by -?-
Calibration on the first calstick cell is complete. Place the transducer on the other calstick cell and start
the CAL2 calibration cycle.
CO₂ ZERO FAILED
Numeric is replaced by -?-
INOP tone
An error occurred during the last zero calibration. Check the airway adapter and clean, if necessary.
Perform another zero calibration. If the INOP persists, contact your service personnel.
CO₂ ZERO REQU'D
Numeric is replaced by -?-
INOP tone
Perform zero calibration for the CO
2
sensor. If the INOP persists, contact your service personnel.
CO₂CHANGE SCALE The CO
2
wave is clipped. Select a more appropriate wave scale to display the whole wave.
CO₂ZERO RUNNING Wait until zero calibration is finished.
INOP Message, Indication What to do
INOP Message, Indication What to do
<SO₂ Label> CAL FAILED
Numeric is replaced by -?-
INOP tone
The calibration failed. Check the catheter-to-Optical-Module connection. Manually restart the
calibration. Try another catheter and Optical Module. If the catheter is already inserted, perform an in-
vivo calibration.
<SO₂ Label> CAL MODE
Numeric is replaced by -?-
INOP tone
Pre-insertion calibration is complete, but the catheter tip is still inside the optical reference. The
catheter is now ready for insertion.
<SO₂ Label> CANNOT
MEAS
Numeric is replaced by -?-
INOP tone
The signal is out of the normal range, and no oxygen saturation can be derived. Perform an in-vivo
calibration. If the INOP persists, try another Optical Module and catheter.
<SO₂ Label> CONN
OPTMOD
Numeric is replaced by -?-
INOP tone
The Optical Module was disconnected during data storage. Reconnect the Optical Module for at least
20 seconds.