3 Patient Alarms and INOPs
82
<SO Label> EQUIP MALF
Numeric is replaced by -?-
INOP tone
The SO
2
/SvO
2
Module or Optical Module is faulty. Unplug and replug the Optical Module and SO
2
/
SvO
2
module. Exchange the modules. If the INOP persists, contact your service personnel.
<SO Label> INCOMPAT.
INOP tone
The SO
2
Module or Optical Module is not supported. Contact your service personnel.
<SO Label> IN-VIVO CAL The in-vivo calibration is not yet complete. Lab values must be stored to the Optical Module to
complete the calibration. Either continue with the next steps of the current calibration or recall the
previous calibration.
<SO Label> LOW LIGHT
Numeric is replaced by -?-
INOP tone
The optical signal levels are too low. Check that the catheter is either in the optical reference or
inserted into the patient. Check the catheter-to-Optical Module connection. If INOP persists, try
another catheter and Optical Module.
<SO Label> NO OPTMOD
Numeric is replaced by -?-
INOP tone
Connect the Optical Module. If the INOP persists, try another Optical Module. Silencing this INOP
switches the measurement off.
<SO Label> OPTMOD
MALF
The Optical Module memory is faulty, and calibration data cannot be stored for transport or during
power failure. If this capability is needed, use another Optical Module.
<SO Label> PRE-INS CAL
Numeric is replaced by -?-
INOP tone
The pre-insertion calibration is running. This typically takes one minute. During this time alarms are
switched off. Wait until the calibration is complete.
<SO Label> UNPLUGGED
Numeric is replaced by -?-
INOP tone
Measurement switched on and SO
2
/SvO
2
module unplugged from the rack.
<SO Label> UPGRADE
INOP tone
The SO
2
module is currently in upgrade mode.
Monitoring is not possible in this mode.
<SO Label> WARMUP
Numeric is replaced by -?-
The Optical Module has not yet reached the operating temperature. Wait a few minutes until warm-up
is finished.
<SO Label>CAL
REQUIRED
Numeric is replaced by -?-
INOP tone
There is no valid calibration data in the Optical Module. Perform either a pre-insertion or an in-vivo
calibration.
<SO Label>CONFIG
ERROR
Numeric is replaced by -?-
INOP tone
The Optical Module has been configured to SaO
2
Mode. Use Change to Venous in the setup
menu to reconfigure to venous saturation mode.
<SO Label>LIGHT INTENS
Numeric is replaced by -?-
INOP tone
The intensity changed considerably since the last light intensity calibration. This may indicate that the
catheter tip is positioned against a blood vessel wall or that there is low blood flow. Reposition the
catheter (and perform a Light Intensity Calibration).