EasyManua.ls Logo

Philips IntelliVue MX800 - Setting up all CO2 Measurements

Philips IntelliVue MX800
460 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13 Monitoring Carbon Dioxide
220
Removing Exhaust Gases from the System
WARNING
Anesthetics: When using the Microstream CO
2
measurement on patients who are receiving or have
recently received anesthetics, connect the outlet to a scavenging system, or to the anesthesia machine/
ventilator, to avoid exposing medical staff to anesthetics.
Use an exhaust tube to remove the sample gas to a scavenging system. Attach it to the MMS Extension
at the outlet connector.
Suppressing Zero Calibration
To prevent an automatic zero calibration from being started in the next five minutes,
•in the
Setup CO menu, select No Zero For 5Min, or select the Suppress CO Zero SmartKey, if
configured.
Selecting
No Zero For 5Min or selecting the Suppress CO Zero SmartKey again before the timer has
timed out resets the timer to five minutes, unless an AutoZero became due during the previous
suppression.
Setting up all CO2 Measurements
These tasks are common to all CO
2
measurements except where otherwise noted.
Adjusting the CO2 Wave Scale
1 In the CO Wave menu or the Setup CO menu, select Scale.
2 Choose a suitable scale range from the pop-up list.
Setting up CO2 Corrections
Temperature, water vapor in the patient's breath, barometric pressure, and the proportions of O
2,
N
2
O
and Helium in the mixture all influence CO
2
absorption. If values seem inaccurately high or low, check
that the monitor is using the appropriate corrections.

Table of Contents

Related product manuals