Philips IntelliVue MX800 - WARNING; Transferring a Centrally-Monitored Patient; 1 Before you disconnect the MMS from the monitor, select the patient name field or select the Admit; Dischrge SmartKey to open the Patient Demographics window, then select the Transfer pop-up key. If the patient is not admitted or not monitored by an...; 2 Remove the MMS and any connected extensions from the monitor.

Philips IntelliVue MX800
460 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 Managing Patients
97
WARNING
If the monitor is not battery-powered, you cannot monitor during transport.
Transferring a Centrally-Monitored Patient
Scenario: A centrally-monitored patient is moved to another monitoring location on the same
Information Center database server without interrupting the collection of patient trend information.
1 Before you disconnect the MMS from the monitor, select the patient name field or select the
Admit/ Dischrge SmartKey to open the Patient Demographics window, then select the Transfer
pop-up key. If the patient is not admitted or not monitored by an Information Center, the
Transfer key is inactive ("grayed-out").
This step preserves the patient's demographic data during the transfer.
2 Remove the MMS and any connected extensions from the monitor.
3 Connect the MMS to the transfer monitor and monitor your patient during the move.
4 At the new location, connect the MMS to the monitor. If the monitor detects a patient mismatch,
a window will open showing your patient's data and asking whether to
Complete Transfer Of
This Patient?
.
5 Select Yes to complete the transfer. This re-admits the patient from the transfer list to the new
monitor. This will upload the patient demographics, and, if configured, the measurement settings
and trend data stored in the MMS to the receiving monitor.
6 Verify that the settings for patient category and paced mode are correct.
Measurement settings for all
measurements (alarm limits,
measurement on/off, etc.)
yes yes, for all MMS and extensions
measurements
Trend data
yes, for all MMS and extensions
measurements (up to a
maximum of 16 or 32,
depending on your database
configuration)
most recent 8 hours of
information, for all MMS and
extensions measurements
Calculation data (HemoCalc
data)
yes no
Events data
yes no
Patient Information Stored in Monitor Stored in MMS and
extensions