36 Specifications
401
Performance Specifications
MX800 Performance Specifications
Power Specifications Power Consumption <200 W average
Line Voltage 100 to 240 V
Current 1.9 to 0.9 A
Frequency 50/60 Hz
Indicators Alarms Off red (crossed out alarms symbol) LED
Alarms red/yellow/light blue (cyan) LED
On/Standby/Error green/red LED integrated in power switch
External Power green LED
Sounds Audible feedback for user input
Prompt tone
QRS tone, or SpO
2
modulation tone
4 different alarm sounds
Remote tone for alarms on other beds in network
Tone for Timer expired
Trends Resolution 12, 16, 24 or 32 numerics @ 12 sec, 1 minute, 5 minute
resolution
Information Multiple choices of number of numerics, resolution and
duration depending on trend option and application area.
For example:
neonatal extended 12 numerics, 24 hours @ 12 secs or
32 numerics 32 hours @ 1 minute
intensive care extended: 16 numerics 120 hours @ 5 minutes
anesthesia extended 32 numerics 9 hours @ 12 seconds
High Res Trend Waves Measurements available HR, SpO
2
, Resp, tcpO
2
, Pulse, Perf, tcpO
2
, CO
2
, AB P, PAP,
CVP, ICP, C PP, BI S, CC O, AWP, A n es t het ic A gen ts, De lt a
SpO
2
, inO
2
Resolution Measurement samples are taken at a resolution of four
samples per second
Update speed waves are drawn at a speed of 3 cm/minute
Events Information trigger condition and time, event classification and associated
detailed view of episode data
Episode data configurable, either:
4 minutes of high resolution trend or
20 minutes of numerics trend @ 12 sec. resolution or
15 seconds of 4 waves @ 125 samples/sec. (Snapshot)
including all current numerics, alarms and inops
Capacity (max) 25 or 50 events for either 8 or 24 hours
Alarm Signal System delay less than 3 seconds
Pause duration 1,2,3 minutes or infinite, depending on configuration
Extended alarm pause 5 or 10 minutes
Review Alarms Information all alarms / inops, main alarms on/off, alarm silence and time
of occurrence
Capacity 300 items

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MX800 and is the answer not in the manual?

Philips IntelliVue MX800 Specifications

General IconGeneral
Resolution1280 x 1024 pixels
TouchscreenYes
Display TypeTFT LCD
Parameters MonitoredECG, SpO2, Temperature
Network ConnectivityEthernet, Wi-Fi
ConnectivityUSB, Ethernet
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C