36 Specifications
423
Inspired Volume
numeric (TVin)
Range 5 to 100 ml (neonatal)
30 to 400 ml (pediatric)
200 to 3000 ml (adult)
Resolution 1 ml (neonatal/pediatric)
10 ml (adult)
Expired Volume
numeric (TVexp)
Range 5 to 100 ml (neonatal)
30 to 400 ml (pediatric)
200 to 3000 ml (adult)
Resolution 1 ml (neonatal/pediatric)
10 ml (adult)
Lung Compliance
numeric
Range 0 to 500 ml/cmH
2
O
Resolution 1 ml/cmH
2
O
Airway Resistance
numeric
Range 0 to 500 cmH
2
O/l/s (neonatal)
0 to 250 cmH
2
O/l/s (pediatric)
0 to 100 cmH
2
O/l/s (adult)
Resolution 1 cmH
2
O/l/s
Peak Inspiratory
Pressure numeric
(PIP)
Range 0 to 120 cmH
2
O
Resolution 1 cmH
2
O
Negative Inspiratory
Pressure numeric
Range -20 to 0 cmH
2
O
Resolution 1 cmH
2
O
Mean Airway
Pressure
Range 0 to 120 cmH
2
O
Resolution 1 cmH
2
O
Plateau Pressure
numeric
Range 0 to 120 cmH
2
O
Resolution 1 cmH
2
O
Positive End-
Expiratory Pressure
numeric (PEEP)
Range 0 to 99 cmH
2
O
Resolution 1 cmH2O
Barometric Pressure
numeric (PB)
Range 400 to 800 mmHg
Resolution 1 mmHg
Peak Inspiratory
Flow numeric (PIF)
Range 0.3 to 25 l/min (neonatal)
1 to 100 l/min (pediatric)
2 to 180 l/min (adult)
Resolution 0.1 l/min (neonatal)
1 l/min (pediatric/adult)
Peak Expiratory
Flow numeric (PEF)
Range 0.3 to 25 l/min (neonatal)
1 to 100 l/min (pediatric)
2 to 180 l/min (adult)
Resolution 0.1 l/min (neonatal)
1 l/min (pediatric/adult)
Inspiration to
Expiration ratio
numeric (I:E)
Range 1: 0.2 ... 9.9
Resolution 0.1
Spirometry Performance Specifications