36 Specifications
406
Analog ECG Output
(ring, tip)
Full scale on display signal gain x measured ECG voltage
Gain error <15 %
Baseline offset <100 mV
Bandwidth 1 to 100 Hz
Output voltage swing ±4 V (min)
Signal delay <22 ms
Signal delay with
older versions of the
M3001A MMS
[identifiable with the
serial number prefix
DE227 or DE441
and option string
#A01]
<30 ms
Pacemaker Pulse filtered and included in ECG output signal
Digital Pulse Output
(ring)
Output low voltage
level
<0.4 V @ I=-1 mA
Output high voltage
level
>2.4 V @ I=1 mA
Pulse Width 100 ms±10 ms (active high)
Pulse Rise Time <1 ms (from 0.4 V to 2.4 V)
Signal delay < 25ms per AAMI EC13
Signal delay with
older versions of the
M3001A MMS
[identifiable with the
serial number prefix
DE227 or DE441
and option string
#A01]
< 35ms per AAMI EC13
X2 (M3002A) Interface Specifications
Measurement Link
(MSL)
Connectors Female ODU (Proprietary)
Power 30 V to 60 V input
Power Sync. RS-422 compliant input 78.125 kHz (typical)
LAN signals IEEE 802.3 10-Base-T compliant
Serial signals RS-422 compliant
Local signals Provided for connecting MMS extensions
MX800 Interface Specifications