36 Specifications
409
Sinus and SV Rhythm Ranges Brady Adult: 15 to 59 bpm
Pedi: 15 to 79 bpm
Neo: 15 to 89 bpm
Normal Adult: 60 to 100 bpm
Pedi: 80 to 160 bpm
Neo: 90 to 180 bpm
Tachy Adult: >100 bpm
Pedi: >160 bpm
Neo: >180 bpm
Bandwidth Diagnostic Mode Adult/neo/pedi: 0.05 to 150 Hz
Extended Monitoring
Mode
Adult/neo/pedi: 0.5 to 150 Hz
Monitoring Mode Adult: 0.5 to 40 Hz
Neo/pedi: 0.5 to 55 Hz
Filter Mode Adult/neo/pedi: 0.5 to 20 Hz
Bandwidth
when the ECG is transmitted from
a telemetry device via short range
radio
Diagnostic Mode Adult/neo/pedi: 0.05 to 40 Hz
Extended Monitoring
Mode
Adult/neo/pedi: 0.5 to 40 Hz
Monitoring Mode Adult: 0.5 to 40 Hz
Neo/pedi: 0.5 to 40 Hz
Filter Mode Adult/neo/pedi: 0.5 to 20 Hz
Differential Input Impedance >2 MΩ RA-LL leads (Resp)
>5 MΩ at all other leads (at 10 Hz including patient
cable)
Common Mode Rejection Ratio Diagnostic mode: >86 dB (with a 51 kΩ/47 nF
imbalance).
Filter mode: >106 dB (with a 51 kΩ/47 nF imbalance).
Electrode Offset Potential Tolerance ±500 mV
Auxiliary Current
(Leads off Detection)
Active electrode: <100 nA
Reference electrode: <900 nA
Input Signal Range ±5 mV
ECG/Arrhythmia/ST/QT
Alarm Specifications
Range Adjustment
HR 15 to 300 bpm
maximum delay: 10 seconds according
to AAMI EC 13-1992 standard
Adult:1 bpm steps (15 to 40 bpm)
5 bpm steps (40 to 300 bpm)
Pedi/Neo:1 bpm steps (15 to 50 bpm)
5 bpm steps (50 to 300 bpm)
Extreme Tachy Difference to high limit 0 to 50 bpm 5 bpm steps
Clamping at 150 to 300 bpm 5 bpm steps
Extreme Brady Difference to low limit 0 to 50 bpm 5 bpm steps
Clamping at 15 to 100 bpm 5 bpm steps
Run PVCs 2 PVCs Not adjustable by user
PVCs Rate 1 to 99 PVCs/minute 1 PVC
Vent Tach HR 20 to 300 bpm 5 bpm
Vent Tach Run 3 to 99 PVCs/minute 1 PVC
ECG/Arrhythmia/ST Performance Specifications