36 Specifications
413
Pulse Range 30 to 300 bpm
Accuracy ±2% or 1 bpm, whichever is greater
Resolution 1 bpm
Sensors Wavelength range: 500 to 1000 nm
Emitted Light Energy: 15 mW
Information about the wavelength range can be especially useful to
clinicians (for instance, when photodynamic therapy is performed)
Pulse Oximeter
Calibration Range
70% to 100%
SpO
2
Alarm
Specifications
Range Adjustment Delay
SpO
2
Adult: 50 to 100%
Pedi/Neo: 30 to 100%
1% steps (0, 1, 2, 3,... 30)
+4 seconds
or Smart Alarm
Delay (see table
below)
Desat Adult: 50 to Low alarm limit
Pedi/Neo: 30 to Low alarm limit
1% steps (0, 1, 2, 3,... 30)
+4 seconds
Pulse 30 to 300 bpm Adult:
1 bpm steps (30 to 40 bpm)
5 bpm steps (40 to 300 bpm)
Pedi/Neo:
1 bpm steps (30 to 50 bpm)
5 bpm steps (50 to 300 bpm)
max. 14 seconds
Tachycardia Difference to high limit 0 to 50 bpm 5 bpm steps max. 14 seconds
Clamping at 150 to 300 bpm 5 bpm steps
Bradycardia Difference to low limit 0 to 50 bpm 5 bpm steps max. 14 seconds
Clamping at 30 to 100 bpm 5 bpm steps
SpO
2
Smart Alarm Delay Specifications
The Smart Alarm Delay functionality is currently not available in the U.S.A. or in clinical environments under FDA control.
Deviation from violated
alarm limit
Resulting alarm delay
Short Medium Long
1% 25 sec
(maximum delay)
50 sec
(maximum delay)
100 sec
(maximum delay)
2% 12 sec 25 sec 50 sec
3% 10 sec 16 sec 33 sec
4% 10 sec 12 sec 25 sec
5% 10 sec 10 sec 20 sec
6% 10 sec 10 sec 16 sec
SpO
2
Performance Specifications