3 Patient Alarms and INOPs
64
** awRR LOW CO
2
, Resp, AGM The airway respiration rate has dropped below
the low alarm limit.
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
** BIS HIGH BIS The Bispectral Index value has exceeded the
high alarm limit.
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
** BIS LOW BIS The Bispectral Index value has dropped below
the low alarm limit.
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
** CCO HIGH
** CCI HIGH
CCO Continuous Cardiac Output or CC Index is
above the high alarm limit.
numeric flashes and high alarm
limit is highlighted, yellow alarm
lamp, alarm tone
** CCO LOW
** CCI LOW
CCO Continuous Cardiac Output or CC Index is
below the low alarm limit.
numeric flashes and low alarm limit
is highlighted, yellow alarm lamp,
alarm tone
** CPP HIGH CPP The CPP value has exceeded the high alarm
limit.
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
** CPP LOW CPP The CPP value has fallen below the low alarm
limit.
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
*** DESAT
*** DESAT xx < yy
SpO
2
The SpO
2
value has fallen below the
desaturation alarm limit. xx denotes the lowest
measured value, and yy is the desaturation limit.
numeric flashes, red alarm lamp,
alarm tone
** EC10 ALARM
*** EC10 ALARM
at Information Center
EC10 Intellibridge A yellow (**) or red (***) patient alarm is
present on the IntelliBridge module. Check the
monitor display for more detailed alarm
information.
(on monitor) yellow or red alarm
lamp; the alarm text is defined by
the Intellibridge device driver
** etCO₂ HIGH CO
2
, Resp, AGM The end tidal CO
2
high alarm limit has been
exceeded.
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
** etCO₂ LOW CO
2
, Resp, AGM The end tidal CO
2
value has fallen below the
low alarm limit.
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
** etO₂ HIGH O
2
, AGM The end tidal O
2
high alarm limit has been
exceeded.
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
** etO₂ LOW O
2
, AGM The end tidal O
2
value has fallen below the low
alarm limit.
numeric flashes, and low limit is
highlighted, yellow alarm lamp,
alarm tone
* EVENT
** EVENT
*** EVENT
at Information center
Event surveillance An event has occurred and the event
notification is configured to alarm. Check on
the monitor for more details on event group.
(on monitor) event group name
flashes, yellow or red alarm lamp
and alarm tone
* EVENT:<Event
Group>
** EVENT:<Event
Group>
*** EVENT:<Event
Group>
Event surveillance An event has occurred and the event
notification is configured to alarm.
event group name flashes, yellow or
red alarm lamp and alarm tone
Alarm Message From Condition Indication