The Patient Session
3-28 ST80i Stress Test System Instructions for Use
To acquire a resting ECG without interpretation
1
Have the patient in a supine position.
2 Click 12 Leads Print button on the Toolbar to obtain a 12-lead resting ECG without
interpretation.
To obtain a 12-lead resting ECG without interpretation:
1
Click the Event button.
2 From the drop-down menu, select Supine.
To acquire a 12-lead standing ECG without interpretation
1
Have the patient in a standing position.
2 Change the lead position to Mason-Likar.
3 Click the Event button.
4 From the drop-down menu, select Standing.
5 Click the 12 Leads Print button on the Toolbar to obtain a 12-lead resting ECG without
interpretation.
After a few seconds, the ST80i system prints a full 12-lead resting ECG with
measurements.
To obtain a 12-lead resting ECG with interpretation
Select an event while in Pre Exercise.
After a few seconds, the ST80i system prints a full 12-lead resting ECG with
measurements and interpretation text.
The next step is to connect the patient to the blood pressure monitor and record a baseline
blood pressure.