Maintaining the ST80i System
5-2 ST80i Stress Test System Instructions for Use
Ordinary equipment
Not suitable for use in the presence of flammable anesthetics
Continuous operation
CAUTION Dispose of waste material according to local regulations and laws regarding medical and battery waste.
Cleaning the ST80i System
To clean the system, including PC, monitor, and transport cart
Clean the mouse every month, as follows:
a Remove the ball from the mouse base.
b Wipe the ball with a dry cloth.
c Reassemble the mouse.
Clean the keyboard with a damp cloth, as necessary.
Clean the outside of the computer with a soft cloth lightly moistened with a mild detergent
solution. Do not use solvents or abrasive cleaners.
Clean the display with a soft cloth lightly moistened with a mild detergent approved for
use on the display. Do not use solvents or abrasive cleaners.
Clean the cart with a damp cloth, as necessary.
ST80i components should
withstand cleaning using a cloth dampened with the following solutions.
IMPORTANT! This applies only to "Philips Healthcare custom-designed components". For all COTS components,
including laser printers, LCD displays, keyboards, and mice, customers must follow the manufacturers'
cleaning specifications.
Isopropyl alcohol, 70% solution
Mild soap and water
Chlorine bleach (5.25% sodium hypochlorite content), 3% solution in water
Steris Coverage Plus NPD, 1/2 fluid ounce per gallon water (one part Coverage Plus NPD
to 256 parts water)
Labels and graphics are to remain intact and legible. There is to be no damage, discoloration,
stress cracking or crazing of parts, cracking or elastomeric swelling. Parts shall not be attacked
to the degree that they lose their gloss or become sticky or gummy.