Working with Reports
4-26 ST80i Stress Test System Instructions for Use
Rhythm Report, Final Report).
Printed reports provide a calibration pulse on printed waveform reports for added check on
print speed and amplitude accuracy.
In the Settings section of the application (Settings System Settings System), you can
preconfigure the printer for each type of report:
Select ECG Report Printer:
–Off
Local Thermal Printer
Select Event Report Printer:
–Off
Local Thermal Printer
Select Rhythm Report Printer:
–Off
Local Thermal Printer
Select Final Report Printer:
–Off
Local Thermal Printer
WARNING ECG signals printed on a laser report should not be used to make time-sensitive or amplitude-sensitive
measurements directly from printed page.
WARNING You are expected to provide power for the laser printer from a source other than the ST80i isolation
transformer in order to be in compliance with your facility's safety requirements or IEC 60601-1.
WARNING The PC, LCD display, thermal printer, desk light, and NIBP/SpO2 module must be plugged into medical
isolation transformer provided with the stress system.
Print Report Settings
During the exercise stress test, ST80i supports several kinds of printing including:
Resting 12- Lead ECG report (with interpretation)
Stage printout (without interpretation)
Event printout (without interpretation)