Working with Reports
4-46 ST80i Stress Test System Instructions for Use
Exporting Reports to an ECG Management System
You can export PDF files for import into the Philips IntelliSpace Enterprise ECG Manager
(version C.04 and later) or the Philips TraceMasterVue ECG Management System (version
C.01 through C.03). You can review the ECG, as well as store it in the database. You cannot,
however, edit the imported PDF file. For details about working with the imported PDF report,
refer to the ECG management system documentation.
If you are exporting to an ECG manager, shared folders with the appropriate permissions must
be set up on the server. You can specify a separate profile for each type of exported file and
also to export to different directories. Check with your system administrator to see which
profile you should use to export to the IntelliSpace or TraceMasterVue systems.
File Naming Conventions
ST80i saves files in a particular naming convention. It is very important that you do not
change this convention and that you enter all required information in the Patient Information
screen. Files that are not named properly will get sent to the TraceMasterVue Error directory.
CAUTION The underscore, square brackets, and lack of spaces are always present in the file name and are not
removable. They are required for proper parsing of the PDF files during import into the
TraceMasterVue database. Stress files must conform to the naming conventions specified here to be
able to be imported into the TraceMasterVue database.