Preparing the Patient
ST80i Stress Test System Instructions for Use 3-35
Controlling the Treadmill or Ergometer
At any point during the stress test, you can switch to “Manual Mode” and change the speed
and elevation grade of the treadmill. This overrides the protocol settings for the current stage.
Once you enter manual mode, there is no next stage.
CAUTION Be sure to inform the patient of the changes.
If the treadmill is unavailable, ST80i displays “Unavailable” where the Start Treadmill
button would be.
To change the speed and elevation of the treadmill
1
Place the pointer in the field that shows the current speed and click.
A box with a minus (-) sign to the left and a plus (+) sign to the right of the current setting
is displayed.
Figure 3-13 Treadmill Spped and Grade Adjustment
2
Make the appropriate adjustment, down or up, and click OK.
3 Place your pointer in the field that shows the current elevation and click.
A box with a minus (-) sign to the left and a plus (+) sign to the right of the current setting
is displayed.
4 Make the appropriate adjustments, down or up, and click OK.
Figure 3-14 “Manual Mode” pop-up warning