ST80i Stress Test System Instructions for Use E-5
Medical Isolation Transformer Specifications
The medical isolation transformers used with the ST80i system meet the following
specifications: UL 60601-1, CSA C22.2 601.1, EN60601.1, CE (LVD), ROHS compliant.
Figure E-1 Medical isolation transformer provided with ST80i
Table E-3 Medical isolation transformer specifications
Function Specifications
Frequency
50/60 Hz
Output ratings
120V 600VA or 240V 600VA
Weight
26.5 lb (12 Kg)
Dimensions
Height.
4.08" (103.5mm)
Width. 6.55" (166.4mm)
Depth. 10.5" (266.7mm)
REF: 1404-102
Input 120VAC – 6.3AT
REF: 1404-003
Input 230VAC – 3.15AT.
Fuses
Use agency-approved fuses only, 250V.
The transformer has an internal thermal resettable fuse.