EasyManua.ls Logo

Philips ST80i - High Ramp Protocol

Philips ST80i
230 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol Reference
B-10 ST80i Stress Test System Instructions for Use
High Ramp Protocol
The Pre-Exercise phase settings are the same as for the Bruce protocol. See Figure X-1 on
page X. The Recovery phase settings are the same as for the Medium Ramp protocol. See
Figure X-8 on page Y.
Table B-10 High Ramp Protocol Settings
High Ramp
Stage # Time
Duration
(mm:ss)
Speed
(mph)
Grade
(%)
ECG Print BP
Measurement
Pre-Exercise
1 Unlimited 1.0 0.0 Of
Off
Exercise
1 0:30 1.6 5.0 End End
2 0:30 1.7 10.0 End End
3 0:30 1.7 10.0 End End
4 0:30 2.0 10.0 End End
5 1.00 2.2 11.0 End End
6 0:30 2.4 11.5 End End
7 0:30 2.5 12.0 End End
8 0:30 2.6 12.5 End End
9 0:30 2.8 13.0 End End
10 1.00 3.0 13.5 End End
11 0:30 3.2 14.0 End End
12 0:30 3.4 14.0 End End
13 0:30 3.5 14.5 End End
14 0:30 3.6 15.0 End End
15 1.00 3.7 15.5 End End
16 0:40 4.0 16.0 End End
17 0:40 4.2 16.0 End End
18 0:40 4.4 16.5 End End
19 0:40 4.6 17.0 End End
20 0:40 4.8 17.5 End End
21 0:40 5.0 18.0 End End
22 0:40 5.2 19.0 End End
23 0:40 5.5 20.0 End End
24 0:40 5.8 21.0 End End
25 0:40 6.0 22.0 End End
Recovery
1 3:00 1.0 0.0 Every
02:00
Every 02:00