Report Phase
ST80i Stress Test System Instructions for Use 4-3
Report Phase
When stress testing enters the Report phase, you can edit the report with the algorithm output
to generate a Final Stress Report. The final report is divided into several pages that include, for
example, the report summary, trends, average QRS, events, and resting ECG.
During the Report phase:
ST80i freezes the Recovery display while the test information is transferred to system
storage.
ST80i loads the Report screen, so you can specify the parameters and format of the final
report.
A real-time 7.5-second ECG is displayed in the lower-left section of the screen for the
Post-Recovery phase
Post-Recovery Phase
A Post-Recovery phase lasts from 5 to 60 minutes. During the Post-Recovery phase, ST80i
will continue to monitor the ECG, arrhythmia alerts, as well as sudden HR changes. Data will
not be saved unless an event is triggered. If events are triggered, the event data is saved as part
of the Final Stress Report.
NOTE ST80i will stop streaming test data after the 60-minutes Post-Recovery phase.