Preparing the Patient
ST80i Stress Test System Instructions for Use 3-39
Report Phase
When you have determined that the patient has completed the Recovery phase of the stress
test, you can manually move to the Report phase where the Report screen is visible.
Data from the Recovery Phase is saved as part of the Final Stress Report. When you save the
stress test, the patient enters the Post-Recovery phase. This phase appears as part of the Report
phase in the Final Stress Report.
Post-Recovery
The Post-Recovery phase begins when the Recovery phase ends, and it stops when the
Report screen is opened. The Post-Recovery phase lasts from 5 to 60 minutes. During the
Post-Recovery phase, ST80i continues to monitor the ECG, arrhythmia alerts, as well as
sudden HR and BP changes. Data will not be saved unless an event is triggered. If events are
triggered, the event data is saved as part of the final report.
See “Notifications and Alerts” on page 3-36
NOTE ST80i will stop stream-test data after the 60-minute Post-Recovery phase. From the Report screen,
you can review and modify the content of the final test report. You can print the individual reports
that are also included in the final report, save the final report, and print it. After you save the report,
you can export it to a PDF file for distribution or storage.
Report Screen
On the Report screen, you can review and edit the stress test results to prepare the Final Stress
Report. The final report includes the Patient Information, Summary, Trends, Average QRS,
Resting ECG results, Events, and Tabular Report. The report is then confirmed, saved,
exported, archived, and/or printed on pre-configured printers. You have the option to print the
whole report or specify which page(s) to be printed.
From the Report screen, you can also review or replay the stress test and finalize (or confirm)
the final report before export.
See “Working with Reports” on page 4-1 for information on the Report Phase and Report
screen.