Electromagnetic Compatibility (EMC)
ST80i Stress Test System Instructions for Use E-9
Guidance and Manufacturer’s Declaration:
Electromagnetic Emissions
The ST80i Stress Test System and optional Thermal Printer are intended for use in the
electromagnetic environment specified in the table below. The customer or the user of the
ST80i Stress Test System and optional Thermal Printer should assure that it is used in such an
environment.
Table E-6 Guidance and Manufacturer’s Declaration: Electromagnetic Emissions
Emissions Test Compliance Electromagnetic Environment Guidance
RF Emissions CISPR 11 Group 1
30 MHz to
1000 MHz
The ST80i Stress Test System and optional Thermal
Printer use RF energy only for its internal function.
Therefore, its RF emissions are very low and not likely
to cause any interference in nearby electronic
equipment.
RF Emissions CISPR 11 Class B
150 kHz to 30 MHz
The ST80i Stress Test System and optional Thermal
Printer are suitable for use in all establishments other
than domestic and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domestic purposes.
Note ST80i PIM and AIM are designed and certified to meet
CFR 47 FCC Part 15 and EN 301 489 for devices with
intentional transmitters.
Harmonic Emissions
IEC 61000-3-2
Complies
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies