An Overview of the ST80i Stress Test System
2-32 ST80i Stress Test System Instructions for Use
Rhythm Print
Before you select the Rhythm Print icon from the Procedure bar to print a continuous strip,
you can modify your settings based on the following:
Leads
Limb Gain
Chest Gain
Speed
To change the Rhythm Print settings
1
Click the Gear (Quick Settings) icon.
2 Select the Rhythm Print tab.
3 Place a checkmark for each lead you want to print. (1 - 13 leads are available.)
4 Use the drop-down arrow to adjust the Limb Gain setting. (2.5, 5, 10, 20 mm/mv)
5 Change the Chest Gain setting:
–Full
–Half
6 Use the drop-down arrow to adjust the Speed. (5, 10, 25, 50 mm/sec)
7 Click OK to save your settings.
Sync Out
ST80i supports two analog ECG output signals and one TTL ECG Sync Output on the
Advanced Interface Module (AIM).
Analog ECG Output - for synchronization with ECHO device for Stress-Echo procedures.
This signal is not to be considered as diagnostic quality.
TTL Sync Output - for QRS gating required by the Tango SunTech implementation of
NIBP measurement
CAUTION Sync ECG output signals are not real-time; therefore, they are not diagnostic quality and should not be
used for analysis.
They are pre-configured as part of the user profile. However, they can be changed during the
patient session.
See the ST80i Installation and Configuration Guide for how to configure Sync Out settings in
the user profile.